Turbidity meter series Turb PLUS 2000
:
- Safe measurements by integrated bubble trap and ultrasonic cleaning
- Easy calibration with standards that are traceable to Formazine
- Standard-compliant measurement according to US EPA 180.1 and DIN EN ISO 7027, respectively
- High accuracy, especially below 40 NTU
The nephelometric turbidity meters of series Turb PLUS 2000 can be used for nearly all applications thanks to its large measuring range. By using white or infrared light and the respective standards, the meters meet the requirements of the US EPA 180.1 as well as those of the DIN EN ISO 7027. Instrument with white as well as infrared light are equipped by default with an integrated bubble trap and ultrasonic cleaning, which saves time and effort. If needed, a set with additional external bubble trap is available.
Calibration solutions need to be ordered separately.
Models:
Turb PLUS 2020: Online turbidity meter, with white light and integrated bubble trap; nephelometric measurement specified
according to US EPA 180.1, with ultrasonic cleaning 110–240 VAC.
Turb PLUS 2120: Online turbidity meter, with infrared light and integrated bubble trap; nephelometric measurement specified according to DIN EN ISO 7027, with ultrasonic cleaning 110–240 VAC.
Turb PLUS 2120 Set: Turb PLUS 2120 including external bubble trap
Specs
Model
Turb PLUS 2020
Turb PLUS 2120
Measurement principle
Scattered light measurement
white light, nephelometric measurementScattered light measurement
infrared light, nephelometric measurementMeasurement Range
Factory setting: 0 ... 100 NTU; optional measuring ranges 0 ... 10 and 0 ... 1000 NTU (loaded on USB)
Resolution
Down to 0.0001 NTU
Accuracy
±2 % of reading or ±0.02 NTU below 40 NTU whichever is greater (0–100 and 0–1000 ranges)
±5 % of reading above 40 NTU (0–100 and 0–1000 NTU ranges)
±2 % of reading or ±0.02 NTU (0–10 NTU range)Limit of Detection
0.02 mg/l
Response time
Adjustable (1 ... 60 readings)
Sample Temperature
1 °C ... 50 °C (34 °F ... 122 °F)
Sample Flow Rate
100 ml/min. ... 1.5 liter/min. (0.026 0.40 Gal/min)
Maximum Water Pressure
Integral pressure regulator rated 7 bar (101.5 PSI)
Cleaning
Ultrasonic cleaning
Calibration
Factory calibrated to 0 100 NTU, recalibration required every three months and after change of measuring range
Display
Multi-Line Liquid Crystal Backlit Display
Mounting
Wall mounting with 6 screws
Regulatory Compliance
Compliant to U.S. EPA 180.1 Compliant to DIN EN ISO 7027
Power
100–240 V AC, 47–63 Hz, 80 VA
Relays
Two Programmable, 120–240 V AC 2A Form C Relay
Analog Output
One powered 4–20 mA, 1000 W drive
Digital Output
Bi-directional RS-485, Modbus RTU
Environmental conditions
Storage Temperature Range: 1 °C ... 50 °C (34 °F 122 °F)
Operating Temperature Range: 1 °C ... 50 °C (34 °F 122 °F)
Relative Humidity: Up to 95 % (non-condensing) Not recommended for outdoor use. Altitude up to 2000 meters.Certification
CE, cETLus
CE Approved, LC listed to UL 61010-1 3rd Edition: 2012 LC: Certified to CSA 22.2 No.61010-1-12: 2012Mechanics
Wetted Materials: Nylon, Wetted Materials: Nylon, Borosilicate Glass, Silicone, Polypropylene, Stainless Steel, Viton, Acety
Housing: ABS (Double Insulated, Pollution Degree 2, Overvoltage Category II)Weight
2.8 kg (6.2 lbs.)